Characteristic Analysis for Dose Area Product Meter and Conversion to Air Kerma
-
- KOSHIDA KICHIRO
- Department of Radiological Technology, School of Health Sciences, Faculty of Medicine, Kanazawa University
-
- SHIMIZU KOUZOU
- Department of Radiological Technology, School of Health Sciences, Faculty of Medicine, Kanazawa University
-
- MIYACHI TOSHIAKI
- Department of Radiological Technology, School of Health Sciences, Faculty of Medicine, Kanazawa University
Bibliographic Information
- Other Title
-
- 面積線量計の特性解析と空気カーマへの検討
Search this article
Description
Among the basic characteristics of the dose area product meter, the following were examined : effect of the heel effect which derives from hardening, atmospheric correction coefficient, response of measuring plane of the DAP meter, target focus angle of an x-ray tube assembly of quality of radiation by the area dosimeter mounting, and conversion from the value of the area dosage to air kerma. In terms of tube voltage, the IEC standard is almost always observed, if the radiation field in the dosimeter's surface area is within 6cm×6cm. Error was 17.6%, even if all factors were not considered, and even if the conversion to air kerma was possible. It was indicated that the conversion to air kerma in the patient was possible when the radiation field and measurement error in the dosimeter's surface were considered.
Journal
-
- Japanese Journal of Radiological Technology
-
Japanese Journal of Radiological Technology 57 (12), 1541-1547, 2001
Japanese Society of Radiological Technology
- Tweet
Keywords
Details 詳細情報について
-
- CRID
- 1390001206390351232
-
- NII Article ID
- 10019610004
- 110003439536
-
- NII Book ID
- AN00197784
-
- ISSN
- 18814883
- 03694305
- http://id.crossref.org/issn/03694305
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- Crossref
- CiNii Articles
- OpenAIRE
-
- Abstract License Flag
- Disallowed