Response Properties of Amperometric CO Sensor Using a Polybenzimidazole (PBI) Membrane above 140 .DEG.C.

DOI Web Site 参考文献14件 オープンアクセス
  • MOCHIZUKI Kei
    Graduate School of Science and Technology, Shizuoka University Gas Equipment Development Center, Yazaki Resources Co., Ltd.
  • IWATSU Hirotaka
    Department of Materials Science and Chemical Engineering, Shizuoka University
  • SUDOH Masao
    Graduate School of Science and Technology, Shizuoka University Department of Materials Science and Chemical Engineering, Shizuoka University
  • ISHIGURO Yoshiaki
    Gas Equipment Development Center, Yazaki Resources Co., Ltd.
  • SUZUKI Takayuki
    Gas Equipment Development Center, Yazaki Resources Co., Ltd.

書誌事項

タイトル別名
  • Response Properties of Amperometric CO Sensor Using a Polybenzimidazole (PBI) Membrane above 140°C

この論文をさがす

説明

A new amperometric sensor based on proton conducting membranes to CO was investigated. It was observed that the sensor current density showed almost linearity against CO concentration in the range of 0–3000 ppm at 200°C. In addition, the sensitivity at less than 300 ppm CO was detected clearly using phosphoric acid doped polybenzimidazole (PBI) membrane. It was found that PBI membrane had high selectivity of CO to H2 in combustor exhaust gases at 200°C. Sensor performance using PBI membrane was less influenced by humidity than Nafion.

収録刊行物

  • Electrochemistry

    Electrochemistry 78 (2), 129-131, 2010

    公益社団法人 電気化学会

参考文献 (14)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ