Prediction of Measurement Precision Based on FUMI Theory for Quantitative Mass Spectrometry with Electron Ionization
-
- TAKAYAMA Mitsuo
- Faculty of Pharmaceutical Sciences, Toho University
-
- HAYASHI Yuzuru
- National Institute of Health Sciences
この論文をさがす
説明
The precision or relative standard deviation (RSD) of measurements in a sector type mass spectrometer equipped with an electron ionization system is examined experimentally and theoretically. The observed RSD is obtained by the usual replication and the theoretical RSD is predicted from the probabilistic properties of the baseline noise by the uncertainty theory called the FUMI theory (Function of Mutual Information). By comparing the experimental and theoretical RSD, this paper identifies two kinds of error sources: (1) for small peaks, the major cause of the measurement uncertainty is the baseline noise originating mainly from the detection unit; (2) for large peaks, the measurement error is subject to ion/particles processes such as the fluctuation of gaseous sample concentration, electron ionization efficiency, and ion beam flux and there is a lower limit of the measurement RSD (≈3%). In case 1, the measurement RSD can successfully be predicted by the theory. In case 2, the observed RSD is even larger than the theoretical RSD, and the repeated measurements are necessary to estimate the measurement uncertainty. The peak corresponding to the limit of detection (RSD=33%) is demonstrated.
収録刊行物
-
- 質量分析
-
質量分析 48 (4), 248-253, 2000
一般社団法人 日本質量分析学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390001206496690688
-
- NII論文ID
- 10012972920
-
- NII書誌ID
- AN0010555X
-
- COI
- 1:CAS:528:DC%2BD3cXntFSntbo%3D
-
- ISSN
- 18804225
- 18843271
- 13408097
-
- NDL書誌ID
- 5432836
-
- 本文言語コード
- en
-
- データソース種別
-
- JaLC
- NDLサーチ
- Crossref
- CiNii Articles
- OpenAIRE
-
- 抄録ライセンスフラグ
- 使用不可