- Other Title
-
- 高真空質量分析計の試作
Search this article
Description
A low background mass spectrometer was constructed. The analyzer tube was made of stainless steel and sealed with copper gaskets, and Alpert type ultra-high vacuum valves were used. The whole apparatus can be baked out at 450°C for many hours. This apparatus has been evacuated for two months and baked out for ten times or so at the temperature of 150-400°C. The final vacuum of1×108mmHg was obtained with the use of ion getter Pump. At that pressure, the residual mass peaks are only m/e=12, 14, 16, 28, and 40. Therefore, this apparatus may be used as a low background mass spectrometer.
Journal
-
- Journal of the Mass Spectrometry Society of Japan
-
Journal of the Mass Spectrometry Society of Japan 1960 (14), 21-27, 1960
The Mass Spectrometry Society of Japan
- Tweet
Details 詳細情報について
-
- CRID
- 1390001206497567488
-
- NII Article ID
- 130002033918
-
- ISSN
- 18804225
- 18843271
- 13408097
-
- Data Source
-
- JaLC
- Crossref
- CiNii Articles
-
- Abstract License Flag
- Disallowed