Spectrophotometric Flow Injection Analysis of Chromium(VI) Coupled with On-line Solid-phase Extraction with Anion-exchange Resin Cartridge Column
書誌事項
- タイトル別名
-
- Original Paper
この論文をさがす
説明
Trace amount of CrO42- in water sample was collected and concentrated on a cartridge type mini-column (4 cm in length and 2 mm i.d.) packed with anion-exchange resin, Toyopearl QAE-550C. The collection and concentration were carried out automatically using a sequential injection procedure. The handling sequence was the steps of cleaning, washing and conditioning of the resin, collection of CrO42- on the resin, and elution of CrO42- from the resin to a detection system. The CrO42- eluate was introduced into the photometric flow injection analysis system using 1,5-diphenylcarbazide as a derivatizing reagent. When an aliquot of 5.0 mL or 20 mL sample solution was treated, the sensitivity increased by 6.4 and 23 times, respectively. By combing the automated on-line pretreatment system with the photometric flow injection system, CrO42- at 10-9 mol L-1 level was determined without expensive detectors such as ICP-AES. On the basis of the interference study, the proposed system was applied to the analysis of a river water sample. Concentration of CrO42- at 2.5x10-9 mol L-1 was determined, and satisfactory results were also obtained with recovery tests.
収録刊行物
-
- FIA研究懇談会会誌
-
FIA研究懇談会会誌 28 (2), 124-, 2011
公益社団法人 日本分析化学会 フローインジェクション分析研究懇談会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390001288049271680
-
- NII論文ID
- 130007430728
-
- ISSN
- 24337374
- 0911775X
-
- 本文言語コード
- en
-
- データソース種別
-
- JaLC
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可