Fundamental Characteristics of the Transient Response of a Heat and Moisture Exchanger with a CaCl<sub>2</sub> Component

Bibliographic Information

Other Title
  • 塩化カルシウムを用いた医療用温湿度交換器の過渡応答特性に関する研究
  • エンカ カルシウム オ モチイタ イリョウヨウオン シツド コウカンキ ノ カト オウトウ トクセイ ニ カンスル ケンキュウ

Search this article

Description

<p>The fundamental characteristics of a medical heat and moisture exchanger with filter (HMEF), the artificial nose, were investigated experimentally and numerically. Transient changes in the temperature and humidity in HMEF were measured. Saturated air at 37°C simulating exhalation by a patient was fed into the HMEF for 180 s, then dry air at 26°C simulating inhalation was fed in the opposite direction. CaCl2 installed in the HMEF intensified the condensation of water and retained water in the HMEF. The water vaporized in the returning dry air, allowing and a patient to inhale air with some humidity. The results also revealed that the condensed water can not be completely returned to the inhalation supply. The numerical results showed the difference in the process of condensation and vaporization.</p>

Journal

  • KAGAKU KOGAKU RONBUNSHU

    KAGAKU KOGAKU RONBUNSHU 44 (6), 355-362, 2018-11-20

    The Society of Chemical Engineers, Japan

References(9)*help

See more

Details 詳細情報について

Report a problem

Back to top