Denture Adhesive Used as Gap Filler when Using a Cooling System Installed Microtome: Effects on Preparation of Thin Serial Sections
-
- Saio Masanao
- Department of Laboratory Sciences, Gunma University Graduate School of Health Sciences
-
- Hatori Mizuho
- Department of Laboratory Sciences, Gunma University Graduate School of Health Sciences
-
- Moriya Syunichi
- Department of Laboratory Sciences, Gunma University Graduate School of Health Sciences
-
- Ogawa Mana
- Department of Laboratory Sciences, Gunma University Graduate School of Health Sciences
-
- Hotoyama Hiroyu
- Department of Laboratory Sciences, Gunma University Graduate School of Health Sciences
-
- Fujimori Misa
- Department of Laboratory Sciences, Gunma University Graduate School of Health Sciences
-
- Nomi Yuri
- Department of Laboratory Sciences, Gunma University Graduate School of Health Sciences
-
- Kobayashi Sayaka
- Department of Laboratory Sciences, Gunma University Graduate School of Health Sciences
-
- Kimura Kiminori
- Department of Hepatology, Tokyo Metropolitan Cancer and Infectious Disease Center, Komagome Hospital
Bibliographic Information
- Other Title
-
- パラフィンブロック冷却装置付きミクロトームと義歯安定剤のブロック裏面隙間充填による連続薄切への効果の検討
- パラフィンブロック レイキャク ソウチ ツキ ミクロトーム ト ギシ アンテイザイ ノ ブロック リメン スキマ ジュウテン ニ ヨル レンゾク ウスギリ エ ノ コウカ ノ ケントウ
Search this article
Description
Objectives: We evaluated the usefulness of a cooling system and the use of denture adhesive (DA) as a gap filler in preparing thin serial sections with a sliding microtome.<br>Methods: We prepared paraffin blocks alone (i.e., no patient sample) and initially examined how DA affected paraffin block cooling by the cooling apparatus. Then, we examined how thin serial sections could be prepared using the cooling system installed in the sliding microtome.<br>Results: With DA, differences in the block surface temperature between 0-5 min and between 0-20 min after cooling were significantly larger than that without DA, although the block temperature (20 min) was not significantly different with or without DA. Additionally, the cooling system installed in the microtome enabled the generation of 1μm-thick serial sections.<br>Conclusions: In daily pathology practice, we typically pre-cool blocks using an ice plate or a cooling equipment, but we do not cool the block during sectioning. This condition sometimes leads to difficulty in preparing serial sections. To solve this problem, a cooling system was installed in the microtome.
Journal
-
- The Kitakanto Medical Journal
-
The Kitakanto Medical Journal 69 (1), 1-6, 2019-02-01
The Kitakanto Medical Society
- Tweet
Keywords
Details 詳細情報について
-
- CRID
- 1390001288142066688
-
- NII Article ID
- 130007626897
-
- NII Book ID
- AN10585677
-
- DOI
- 10.2974/kmj.69.1
-
- ISSN
- 18811191
- 13432826
-
- HANDLE
- 10087/00014000
-
- NDL BIB ID
- 029536954
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- IRDB
- NDL
- Crossref
- CiNii Articles
-
- Abstract License Flag
- Disallowed