CT 検査における位置決め撮影の被ばく線量

書誌事項

タイトル別名
  • Radiation Exposure in Computed Tomography Localizer Radiograph
  • CT ケンサ ニ オケル イチギメ サツエイ ノ ヒバクセンリョウ

この論文をさがす

説明

<p>This research measured the radiation exposure of the computed tomography(CT) localizer radiograph of the trunk of the body. The entrance surface dose for CT localizer radiograph was measured using radiophotoluminescent glass dosimeter(RPLD) on four points of measurement, including the center of the phantom, on the surface of a phantom placed in the center of a CT bed, assuming that the subject has a thickness of 20 cm. The entrance surface dose of the localizer radiograph under the chest CT protocol manufacturer’s initial setting conditions of 120 kV 35 mA was 0.80 mGy at the center and 0.53 for the 4-location average for the upper X-ray tube (excluding the CT bed), and 0.74 mGy at the center and 0.48 mGy for the 4-location average for the lower X-ray tube (including the CT bed). Compared to the Japan DRLs 2015 chest X-ray (P→A), the entrance surface dose was 2.67 times at the center and 1.77 times for the 4-location average for the upper X-ray tube and 2.47 times at the center and 1.60 times for the 4-location average for the lower X-ray tube. The CT radiation dose also cannot be ignored for the localizer radiograph entrance surface dose.</p>