A Contact-less Heartbeat Monitoring System for In-Vehicle RRI Measurement
-
- Igawa Nao
- Okayama Prefectural University
-
- Yokogawa Tomoyuki
- Okayama Prefectural University
-
- Fujii Kento
- Okayama Prefectural University
-
- Kayano Isao
- Kawasaki University of Medical Welfare
-
- Sato Yoichiro
- Okayama Prefectural University
-
- Arimoto Kazutami
- Okayama Prefectural University
Bibliographic Information
- Other Title
-
- 車内におけるRRI計測を目的とした非接触心拍モニタリングシステムの検討
- シャナイ ニ オケル RRI ケイソク オ モクテキ ト シタ ヒセッショク シンパク モニタリング システム ノ ケントウ
Search this article
Abstract
<p>In this paper, we propose a contact-less heartbeat monitoring system for in-vehicle R-R Interval (RRI) measurement using a UWB sensor. Our system extracts a heartbeat by noise filtering with a bandpass Butterworth filter to remove a noise caused by inside/outside of the driving vehicle. We carried out an experiment to evaluate the accuracy of RRI detection of our system by comparing it with MIMO radar, which is used widely for moving object detection. We evaluated the accuracy of RRI detection of our system by using Bland-Altman plot with the RRI obtained by myBeat, which is an existing contact-type heartbeat monitoring device. Consequently, the RRI measured by our system is more accurate than that measured by MIMO radar and is comparable to the RRI measured by myBeat.</p>
Journal
-
- IEEJ Transactions on Sensors and Micromachines
-
IEEJ Transactions on Sensors and Micromachines 140 (10), 256-264, 2020-10-01
The Institute of Electrical Engineers of Japan
- Tweet
Keywords
Details 詳細情報について
-
- CRID
- 1390004222630202368
-
- NII Article ID
- 130007920659
-
- NII Book ID
- AN1052634X
-
- ISSN
- 13475525
- 13418939
-
- NDL BIB ID
- 030688619
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- Abstract License Flag
- Disallowed