Optimal Magnification Ratio of Direct Macroradiography in High Magnification Modulation Transfer Function Study on System Combined with Intensifying Screen-Film and Object
この論文をさがす
説明
Calculation of the composite modulation transfer function (MTF) of macroradiography in high magnification conducted with an X-ray tube having a very fine focal spot of 50 μ was performed with the purpose of obtaining the optimal magnification ratio. The optimal magnification ratio of a macroradiogram taken without the object of scatterer was 4-fold when a Fuji KX medical X-ray film combined with fine definition intensifying screen of Kyokko FS was used, 5-fold with medium speed screen of Kyokko MS and 6-fold with high speed screen of Kyokko HS. The optimal magnification ratio was 8-fold regardless of the kind of intensifying screen-film system when the experiment was made with a water phantom of 5-15 cm in thickness as scatterer. It was indicated from MTFs of macroangiograms of various magnifications obtained in animal experiments that the image quality of the macroangiogram of 6- or 8-fold magnification was superior to those of 2-, 4-, 10- and 15-fold ones. It was concluded from the examination of macroangiograms of animals (the part of the rabbit examined was 3-5 cm thick) and macrolymphograms of patients (the examined part was 13 cm thick) that the macroradiogram of 6- or 8-fold magnification produces better findings of the fine structure of vessels than macroradiogram of other magnifications. The results accorded well with those obtained by phantom experiments with water of 5-15 cm in thickness as scatterer.
収録刊行物
-
- Nagoya Journal of Medical Science
-
Nagoya Journal of Medical Science 34 (3), 227-240, 1972-02
Nagoya University School of Medicine
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390009224622682624
-
- NII論文ID
- 120005333568
-
- ISSN
- 21863326
- 00277622
-
- HANDLE
- 2237/18696
-
- 本文言語コード
- en
-
- 資料種別
- departmental bulletin paper
-
- データソース種別
-
- JaLC
- IRDB
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用可