畜産草地研究所大中家畜代謝実験棟における新データ収集処理装置の設計

書誌事項

タイトル別名
  • new data acquisition system for respiration trial system on metabolism laboratory in National Institute of Livestock and Grassland Science
  • チクサン ソウチ ケンキュウジョダイチュウ カチク タイシャ ジッケントウ ニ オケル シン データ シュウシュウ ショリ ソウチ ノ セッケイ

この論文をさがす

説明

A new data acquisition system for open-circuit indirect calorimetry has been constructed and validated. Equipments of the system are consisting with commercially-available computer and input-output control units. The function of the software is specialized to monitor and calculate the respiration trial data under personal computer system, and the design of the software is oriented to simplicity, flexibility and user-friendliness. When the respiration trial data acquired from this system were compared with the data from existing system, the relative errors were very low (0.1~0.7 ± 0.1~0.3%), the correlations were very high (r = 0.99562~0.99997), and few systematic errors were found. The data obtained from this system were sufficient, thus, it is assumed that the function of new system is equivalent to the existing system. Furthermore, the new system has flexibility for expansion or reformation of apparatus in the future use.

大中家畜代謝実験棟の呼吸試験装置について, データ収集処理装置を再構築し, その測定値を従来のものと比較検討した。データ収集処理装置はできるだけ一般的に入手可能な機器で構成し, 将来の改良や更新に十分対応できるように設計した。また, ソフトウェアは簡便でかつ発展性が高くなるよう設計した。すなわち, 1. 一般的に利用されているオペレーティングシステム上で動作し, 起動および操作が簡単に行えるようにした。2. 呼吸試験データの収集と演算に機能を特化し, 簡素な構成にした。3. 設備や機器構成の変更にも柔軟に対応できるような仕様でプログラミングを行った。今回設計したデータ収集処理装置と従来のものとで同時に呼吸試験データを収集してデータを比較したところ, 相対誤差 (0.1~0.7±0.1~0.3%) は小さく, 相関 (相関係数 r=0.99562~0.99999) は高く, 加えて両システム間の系統誤差は軽微であったことから, 今回設計したデータ収集処理装置は, 従来のものと同等の呼吸試験データが得られる非常に簡便なシステムであるといえた。