Interference-Induced Voltage Estimation of Implantable Medical Devices near Electric and Magnetic Field Antennas in Sub-6 GHz / 5G Frequency Band

DOI HANDLE Open Access

Bibliographic Information

Other Title
  • 第5世代移動通信システムSub-6 GHz周波数帯における電界型及び磁界型アンテナ近傍での植込み型医療機器干渉誘起電圧推定

Abstract

This paper describes a numerical assessment of implantable medical devices (implantable cardiac pacemaker/implantable defibrillator) electromagnetic interference (EMI) triggered by wireless devices. By using three-dimensional full-wave numerical simulation based on finite element method, interference voltage induced at the connector of pacemaker inside torso phantom that is used for in-vitro EMI assessment is obtained. We discuss EMI characteristics due to irradiation by slot antenna comparing with irradiation by a dipole antenna. Furthermore, in order to validate the simulated results, we achieved measurement of interference voltage induced at pacemaker connector inside test phantom using novel electro-optic sensor installed in a pacemaker mock-up.