単孔式腹腔鏡手術用鉗子のための腹壁迂回導管部の開発

書誌事項

タイトル別名
  • Bypass Mechanism Arm for Single Incisional Laparoscopic Surgery Forceps

抄録

<p> Single Incisional Laparoscopic Surgery (SILS) is executed through a single 2 cm incision in the navel, it makes forceps touch each other, and doesn't homologize a movement between operational parts and tip forceps parts.</p><p> In our research, we develop new type of laparoscopic surgery forceps, "Bypass forceps". Bypass forceps is inserted through navel but brings the operational direction in line with forceps grasping direction by parallel linkage.</p><p> We have done designing the links of Bypass Mechanism Arm (BMA), simulating and evaluating motion of BMA, fabricating and verifying prototype forceps. The operation direction of the forceps could not be aligned with the gripping direction. The desired posture and movement were not obtained, and the parallel link mechanism did not become parallel exactly.</p><p> As a result, we confirmed that a processing method with a high degree of parallelism was required.</p>

収録刊行物

参考文献 (9)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ