Comparison of ventilator display values and measured device values for tidal volume in volume-controlled ventilation using an exhalation port type single breathing circuit

  • Hamasaka Shuichi
    国立病院機構医王病院 呼吸器障害・感染症治療研究室 金沢大学大学院 医薬保健学総合研究科保健学専攻
  • Tagami Atsuro
    国立病院機構医王病院 呼吸器障害・感染症治療研究室

Bibliographic Information

Other Title
  • 呼気ポート式シングル呼吸回路を用いた従量式換気における機器表示値と測定値の一回換気量比較分析

Description

<p>In this study, our aim is to quantitatively evaluate and analyze the ventilator display values and the measured device values for tidal volume using three types of respiratory mechanics, for the purpose of conducting a comparative analysis. The artificial ventilators targeted in this study are the Philips Trilogy Evo series and the Breas Medical AB mart Ventilator Vivo 45LS, both of which utilize an exhalation port type single breathing circuit. Measurements were conducted using a measuring instrument and a test lung with adjustable compliance and resistance. As a result, a significant difference was found between the ventilator display values and the measured device values for tidal volume, regardless of the test lung settings. When setting the tidal volume of a ventilator, it is important not to rely exclusively on the displayed values.</p>

Journal

Citations (1)*help

See more

References(3)*help

See more

Details 詳細情報について

Report a problem

Back to top