Determination of tellurium in lake water by ICP-MS.

  • FUJINO Osamu
    Research Institute for Science and Technology, Kinki University
  • HARA Kunita
    Department of Chemistry, Faculty of Science Technology, Kinki University
  • IKEJIMA Satoru
    Department of Chemistry, Faculty of Science Technology, Kinki University
  • GODA Shiro
    Department of Chemistry, Faculty of Science Technology, Kinki University

Bibliographic Information

Other Title
  • 誘導結合プラズマ質量分析法による湖水中テルルの定量
  • ユウドウ ケツゴウ プラズマ シツリョウ ブンセキホウ ニヨル コ スイチュウ

Search this article

Abstract

Inductively coupled plasma-mass spectrometry (ICP-MS) was applied to determine the tellurium in lake water. Although ICP-MS is highly sensitive (ppb) to tellurium, the direct analysis of lake water is very difficult because of the extremely low concentration (ppt). Further, the intensity of tellurium was changed due to coexisting barium. It was therefore necessary to concentrate tellrium and to separate it from barium. After the tellurium in lake water was concentrated and separated from barium by coprecipitation with ferric hydroxide, the precipitate was dissolved in hydrochloric acid. The ferric ion was removed by a solvent-extraction method with HPMTFP-butyl ether system, and the aqueous phase was directly injected into the ICP-MS apparatus. Tellurium in Lake Biwa was found to be 0.0360.043 ppb.

Journal

  • BUNSEKI KAGAKU

    BUNSEKI KAGAKU 46 (11), 857-862, 1997

    The Japan Society for Analytical Chemistry

Citations (3)*help

See more

References(9)*help

See more