水素化物原子吸光法による大気粉じん中のアンチモンの定量

書誌事項

タイトル別名
  • Determination of antimony in ambient particulates by hydride generation atomic absorption spectrometry.
  • スイソカブツ ゲンシ キュウコウホウ ニ ヨル タイキ フンジンチュウ ノ ア

この論文をさがす

抄録

Antimony in ambient particulates was digested with HF/HNO3/H2SO4/KMnO4, and determined by hydride generation atomic absorption spectrometry. The recommended procedure is as follows: 0.050.1 g of particulate sample, 15 ml of nitric acid, 1 ml of sulfuric acid, 10 ml of 46% hydrofluoric acid and 2 ml of 2% potassium permanganate solution were taken in a teflon beaker and heated for 1 h on a hot plate at 250°C. Then the solution was evaporated to nearly dryness. After cooling, the residue was dissolved with 20 ml of 6 M hydrochloric acid, and the solution was diluted to 50 ml with water. Twenty milliliter of this solution (containing less than 0.3 μg of antimony) was transfered to a reaction vessel, then 4 ml of hydrochloric acid, 1 ml of 40% potassium iodate solution and 1 ml of 20% L-ascorbic acid solution were added, and the reaction vessel was attached to a hydride generation apparatus. After the air in the reaction vessel was flushed out with nitrogen stream, 7 ml of 1% sodium borohydride solution was injected. The generated stibine was collected in the vessel for 30 s, and then the generated gas was introduced by nitrogen stream into the quartz cell heated at 1000°C. The proposed method was also applied for the determination of antimony in Urban Particulate Matter of NBS SRM-1648.

収録刊行物

  • 分析化学

    分析化学 34 (10), 646-648, 1985

    公益社団法人 日本分析化学会

被引用文献 (2)*注記