High-performance liquid chromatography of amino acids with on-column fluorescence derivatization for a neonatal mass-screening test.
-
- YONEDA Yutaka
- Toyama Institute of Health
-
- TAKAHASHI Masako
- Toyama Institute of Health
-
- KITAMURA Takashi
- Toyama Institute of Health
Bibliographic Information
- Other Title
-
- オンカラム誘導体化高速液体クロマトグラフィーによる新生児マススクリーニング血液ろ紙中のアミノ酸の定量
- オン カラム ユウドウタイカ コウソク エキタイ クロマトグラフィー ニヨル
- オンカラム誘導体化高速液体クロマトグラフィーによる新生児マススクリーニング血液濾紙中のアミノ酸の定量
Search this article
Abstract
A determination method of amino acids in a dried blood disc for a neonatal mass-screening test was developed. The method consists of extraction with a 4mM sodium borate aqueous solution from a dried blood disc treated with methanol-acetone (1: 1) and an on-column fluorometric derivatization. High-performance liquid chromatography was performed isocratically using an acetonitrile-4 mM sodium borate aqueous solution (14.6: 85.4, v/v) containing 2 mM o-phthalaldehyde and 2 mM N-acetyl-L-cysteine as a mobile phase. Amino acids (i.e. Tyr, Val, Met, Ile, Orn, Phe, Leu, Lys) and methylhistamine as an internal standard were eluted within fifteen minuits with excellent reproducibility.