種々の温度における水-アルコール類混合溶媒系中の鉛イオン選択性電極の応答性

DOI 機関リポジトリ NDLデジタルコレクション HANDLE Web Site ほか1件をすべて表示 一部だけ表示

書誌事項

タイトル別名
  • 種々の温度における水‐アルコール類混合溶媒系中の鉛イオン選択性電極の応答性
  • Response of a lead ion selective electrode in water-alcohol systems at different temperatures
  • シュジュ ノ オンド ニ オケル ミズ アルコールルイ コンゴウ ヨウバイケイ

この論文をさがす

抄録

The response of an Orion Lead Electrode 94-82A at 10.0 °C, 25.0 °C, 35.0 °C, 45.0 °C and 50.0 °C(± 0.05 °C) was examined in aqueous 0.1 mol dm-3 KNO3 systems containing alcohols (methanol, ethanol, 1-propanol, ethylene glycol and glycerol). The results were conveniently analyzed by means of a new type of plot in which (2F/2.303 RT) x (E-E_??_) is plotted against log([Pb2+]/c_??_), where E is the potential of the electrode when lead ion concentration is [Pb2+], E_??_ is the potential observed ([Pb2+]=c_??_), and the other symbols have their usual meanings. The advantage of this plot is that it gives a straight line with a slope of 1 passing through the origin when the response is nernstian irrespective of temperature and junction potential, if c_??_ is rightly chosen. In the present case the value of c_??_ was 10-3 mol dm-3. At lead ion concentrations ranging between 1 and 10 mmol dm-3 the response was nernstian in every case. At concentrations lower than a certain limit, which was around (10-410-5) mol dm-3 in the majority of cases, the magnitude |E-E_??_| became smaller than the ideal response. Some samples of 1-propanol contained unknown impurities that interfered with the response of the electrode.

収録刊行物

  • 分析化学

    分析化学 29 (10), 659-663, 1980

    公益社団法人 日本分析化学会

詳細情報 詳細情報について

問題の指摘

ページトップへ