マルチコレクター誘導結合プラズマ質量分析法による高精度同位体分析のための同位体差別効果の補正法(<特集>質量分析法の応用技術)

書誌事項

タイトル別名
  • Evaluation of correction method for mass discrimination effect in multiple collector inductively coupled plasma mass spectrometry.
  • マルチコレクター誘導結合プラズマ質量分析法による高精度同位体分析のための同位体差別効果の補正法
  • マルチコレクターユウドウ ケツゴウ プラズマ シツリョウ ブンセキホウ ニヨル

この論文をさがす

抄録

This paper describes advances in isotopic measurements that have been made with an inductively coupled plasma source magnetic sector multiple collector mass spectrometer (MC-ICP-MS) and presents results of new experiments aimed at further evaluating the instrumental capability as well as the correction technique for the mass discrimination effects. The ability to correct for the mass discrimination effect using a second element of similar mass and very high sensitivity for elements that are otherwise difficult to ionize gives this instrument major advantages over other conventional techniques for isotopic measurements.<BR>The isotopic data obtained by MC-ICP-MS clearly demonstrate potential as a new technique to produce precise and reproducible isotopic data for the elements that are difficult to measure by thermal ionization mass spectrometry (TIMS).

収録刊行物

  • 分析化学

    分析化学 45 (6), 465-472, 1996

    公益社団法人 日本分析化学会

被引用文献 (14)*注記

もっと見る

参考文献 (9)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘