書誌事項
- タイトル別名
-
- Relationship between Asymmetrical Position of Electrode and the Sensitivity Coefficient of a Triode Type Ionization Gauge.
- サンキョクカンケイ デンリ シンクウケイ ノ デンキョク ヘンイ ト カンド ケイスウ ノ カンケイ
この論文をさがす
抄録
Secondary standard ionization vacuum gauge VS-1 is a triode gauge. Simulation of the electrical potential profile of the gauge, electron trajectory emitted from a hot filament, and ion current produced by ionization of gas by the electron assuming a simple electrode geometry was done. The increase in sensitivity coefficient depending on grid potential and the decrease in that depending on filament potential agreed with experimental results by 5%. The increase in sensitivity coefficient depending on filament displacement and ion collector tilting from its regular position was also explained by simulation.