VAMAS‐SCA‐WG in JAPAN活動報告

DOI Web Site 被引用文献1件 オープンアクセス

書誌事項

タイトル別名
  • Report of VAMAS-SCA-WG in Japan.
  • VAMAS SCA WG in Japan カツドウ ホウコク

この論文をさがす

説明

Under the auspices of the 141 comittee (Microbeam Analysis) of Japan Society for Promotion of Science, the AES working group has commenced its work on 1) the establishment of the measurement conditions and 2) the comparison of standard spectra obtained by various spectrometers. Attention was particularly paid to taking the Auger signals at an optimum position of the specimen with respect to the energy analyzer. Systematic investigations on the procedures for energy analyzers, cylindrical mirror analyzers (CMA), are reported in the first section of the present paper. Three kinds of calibration procedures were proposed for attaining the optimum positioning of the specimen with respect to the CMA. Elastic peak varied sensitively in intensity and peak shape, around the optimum position of the specimen.<BR>The comparison of AES instruments installed at several institutes was carried out by measurement of the derivative Auger peak intensity ratio for pure gold and copper under a specific measurement condition, proposed in the first section. The results are shown by the Youden plot following the laborious work performed by the ASTM comittee E-42.