The Difference of the Pattern Coefficient of the Quadrupole Mass Spectrometer.
-
- NAGAYAMA Norihito
- Faculty of Engineering, Osaka City University
-
- ARAKI Kazuhiko
- Faculty of Engineering, Osaka City University
-
- MIMA Hiroshi
- Faculty of Engineering, Osaka City University
Bibliographic Information
- Other Title
-
- 四重極型質量分析計のパターン係数のイオン源による違い
- 4ジュウキョクガタ シツリョウ ブンセキケイ ノ パターン ケイスウ ノ イオ
Search this article
Abstract
A new method is proposed for the derivation of mass spectrum cracking pattern which is supposed to be possessed by ions in a variety of electron bombardment type ion sources. The cracking pattern of Argon in grid type, axial type and Nier type ion source is obtaind. The pattern coefficient was derived from partial ionization cross section of the sample gas with aid of l-V distribution data of ionizing electron, which represents the specific ionizing electron path length li of an electron having an energy of eVi.
Journal
-
- Shinku
-
Shinku 40 (8), 650-654, 1997
The Vacuum Society of Japan
- Tweet
Keywords
Details 詳細情報について
-