-
- 齊藤 一朗
- 電子技術総合研究所
書誌事項
- タイトル別名
-
- Establishment of Absolute Spectral Response Standard at 633 nm
- ハチョウ 633nm ニ オケル ゼッタイ ブンコウ オウトウド ヒョウジュン
この論文をさがす
説明
Electrotechnical Laboratory has established a standard of spectral response at 633 nm based on the self-calibration of a silicon photodiode and an absolute radiometer (operating at room temperature). The standard was maintained a windowed silicon photodiode and transferred by a calibration system using a monochromator. The change in the response of the windowed silicon photodiode was 0.04% over five years. The accuracy of this standard was verified through three international comparisons: the first one organized by CCPR in 1986, the second by CIE in 1988, and the third by CCPR in 1992. The uncertainty of the standard is estimated to be 0.2%.
収録刊行物
-
- 照明学会誌
-
照明学会誌 78 (10), 568-573, 1994
一般社団法人 照明学会
- Tweet
キーワード
詳細情報 詳細情報について
-
- CRID
- 1390282679061252608
-
- NII論文ID
- 130006764559
- 40004320606
-
- NII書誌ID
- AN00268860
-
- ISSN
- 1349838X
- 21851506
- 00192341
-
- NDL書誌ID
- 3891936
-
- データソース種別
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可