全身用ポジトロンCT装置POSITOLOGICA IIの計数率特性と数え落としの補正

書誌事項

タイトル別名
  • Count rate characteristics and count loss correction of Positologica II: A whole body positron emission tomograph.

抄録

This paper describes evaluation and correction of count rate characteristics of POSITOLOGICA II, a multi-slice whole body positron emission tomography system. The present study was performed using three phantoms; (1) a 5 cm inner diameter, water-filled lucite cylinder, (2) a 20 cm inner diameter, water-filled lucite cylinder and (3) a chest phantom. After injection of high activity (about 1.85 GBq (50 mCi) ) of 13N ammonia into each phantom, rates of true coincidence, random coincidence and single photon detections were measured during decay of the isotope through more than two orders of magnitude of activity. At very high levels of activity, count rate characteristics of the system were saturated and limited to 660 kcps of total coincidence rate, which was the sum of rates in on-time and off-time windows, by the FIFO (first-in first-out) output frequency. Below those levels of activity the relationship between count loss and true coincidence rate was not unique but depended on the phantom configurations, suggesting that count loss correction using the above relationship was inadequate for quantitative study. However, the relationship between count loss and single rate was almost independent of the phantom configurations. Thus in conclusion count loss could be corrected using single rate for POSITOLOGICA II. A practical method of count loss correction was also proposed.