Energy Response of a Skin Dose Monitor (SDM model 104-101).
-
- UCHIDA Isao
- Department of Radiation Protection, Cancer Institute Hospital
-
- MIURA Tadashi
- Department of Radiology, Tokyo College of Electronics
-
- TAKAHASHI Tamio
- Department of Radiotherapy, Cancer Institute Hospital
-
- KUWAHARA Akio
- Department of Radiotherapy, Cancer Institute Hospital
-
- ITO Akira
- Department of Physics, Cancer Institute
-
- YAMASHITA Takashi
- Department of Radiotherapy, Cancer Institute Hospital
Bibliographic Information
- Other Title
-
- 積算皮膚線量計(SDM‐モデル104‐101)のエネルギー特性
- セキサン ヒフ センリョウケイ SDM-モデル 104-101 ノ エネルギー
Search this article
Description
The purpose of this study is to estimate the energy response of a SKIN DOSE MONITOR (SDM model 104-101), which was developed to measure and assess medical exposure conveniently. The energy responses of pre-calibrated sensors were estimated by comparative measurements in the same calibration fields with sensors calibrated by Japan Quality Assurance Organization. The relative energy responses obtained for the precalibrated sensors agreed approximately with those of the calibrated sensors. The differences of the calibration factors in different calibration fields for the pre-calibrated sensors were 5 percent at the most in the range from 30keVto 40keVof effective energies. The sensor and coupling optical fiber of the SDM are faintly visible on a radiograph taken under ordinary operating conditions. The influence on diagnosis of this phenomenon is to be considered hereafter.
Journal
-
- RADIOISOTOPES
-
RADIOISOTOPES 47 (10), 759-764, 1998
Japan Radioisotope Association
- Tweet
Details 詳細情報について
-
- CRID
- 1390282679130489728
-
- NII Article ID
- 130004127469
- 10002683771
-
- NII Book ID
- AN00351589
-
- ISSN
- 18844111
- 00338303
-
- NDL BIB ID
- 4583798
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- Abstract License Flag
- Disallowed