Influence of Unit Voxel Size of Cumulated Radioactivity Distribution on 3D Absorbed Dose Estimates.

Bibliographic Information

Other Title
  • 三次元吸収線量評価における累積放射能分布のボクセルサイズの影響
  • 3ジゲン キュウシュウ センリョウ ヒョウカ ニ オケル ルイセキ ホウシャノ

Search this article

Abstract

In radiotherapy due to administration of radionuclides, accurate estimates of three-dimensional (3D) absorbed dose distributions of target regions can be performed using a 3D dose matrix convolution method which has already been developed based on a voxel based calculation algorithm by the authors.<BR>The purpose of this study is to elucidate the influence of unit voxel sizes of cumulated radioactivity distributions of beta-emitting radionuclides on 3D absorbed dose estimates.<BR>Computer simulations were performed using cubic phantoms (24×24×24 mm, 32×32×32 mm, and 48×48×48 mm) and the MIRD thyroid phantom as the target regions assumed distributed uniformly for various beta-emitting radionuclides (32P, 90Y, 131I, 186Re, and 188Re) . The maximum, mean and minimum absorbed doses for their phantoms comprised of various unit voxel sizes (1×1×1 mm, 4×4×4 mm, and 8×8×8 mm) were estimated using the 3D dose matrix convolution method developed.<BR>In the cases of the 3D cumulated radioactivity distributions of the target regions comprised of various unit voxel sizes were identical with one another, the maximum and mean absorbed doses for the target volumes were independent of the unit voxel sizes. However, as the unit voxel sizes were increased, the minimum absorbed doses were overestimated.

Journal

  • RADIOISOTOPES

    RADIOISOTOPES 46 (3), 154-160, 1997

    Japan Radioisotope Association

References(11)*help

See more

Details 詳細情報について

Report a problem

Back to top