Molecular Weight Determination of Hypromellose Acetate Succinate (HPMCAS) Using Size Chromatography with a Muiti-Angle Laser Light Scattering Detector

  • Fukasawa Miyuki
    Specialty Chemicals Research Center, Shin-Etsu Chemical Co., Ltd.
  • Obara Sakaé
    Specialty Chemicals Research Center, Shin-Etsu Chemical Co., Ltd.

書誌事項

タイトル別名
  • Molecular Weight Determination of Hypromellose Acetate Succinate (HPMCAS) Using Size Exclusion Chromatography with a Multi-Angle Laser Light Scattering Detector

この論文をさがす

説明

The molecular weight of hypromellose acetate succinate (HPMCAS), a polymer used for enteric coating, was determined by means of size exclusion chromatography with a multi-angle laser light scattering detector. The weight-average molecular weight (Mw) of several lots and grades ranged approximately from 17000 to 20000, and the number-average molecular weight (Mn) was around 13000. The inter-day precision of measurement, in terms of the coefficient of variation, was less than 5%.

収録刊行物

参考文献 (4)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ