Determination of Arsenic by High Power Nitrogen Microwave Induced Plasma Atomic Emission Spectrometry with Hydride Generation Technique.

書誌事項

タイトル別名
  • Determination of Arsenic by High Power

この論文をさがす

抄録

High power nitrogen microwave induced plasma was used as a new excitation source for atomic emission spectrometry to determine low concentrations of arsenic in combination with hydride generation technique. Under the optimized experimental conditions, the best attainable detection limits at As(I)188.979nm was 3.13ngAs/ml with a linear dynamic range of 10 to 5000ngAs/ml. The presence of several diverse elements has been found to cause more or less a depressing interference with the determination of arsenic by the present technique. The present method was applied to the determination of total arsenic in the samples of sea weed and scallop to be issued as standard reference materials by National Institute for Environmental Studies, Japan. The samples were dissolved in a mixture of nitric and sulfuric acids with the addition of hydrofluoric acid. The results obtained by this method were very reproducible and satisfactory.

収録刊行物

  • Analytical Sciences

    Analytical Sciences 13 (Supplement), 13-16, 1997

    社団法人 日本分析化学会

被引用文献 (1)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ