Multielement Analysis of Suspended Particulates Collected with a Beta-Gauge Monitoring System by ICP Atomic Emission Spectrometry and Mass Spectrometry.

DOI PDF Web Site Web Site Web Site View 1 Remaining Hide 4 Citations 30 References
  • WANG C. F.
    Department of Nuclear Science, National Tsing Hua University
  • CHIN C. J.
    Department of Nuclear Science, National Tsing Hua University
  • CHIANG Pen-Chi
    Institute of Environmental Engineering, National Taiwan University

Bibliographic Information

Other Title
  • Multielement Analysis of Suspended Part

Search this article

Abstract

This work presents a two-step HNO3+HClO4/HF(1:2:1, v/v) acid mixture microwave digestion method for determining the elemental concentrations in suspended particulates collected on glass fiber filters by a beta-gauge monitoring system. Eighteen elements (Al, As, Ba, Ca, Cd, Cr, Cu, Fe, K, Mg, Mn, Na, Ni, Pb, S, Sb, V and Zn) were determined using inductively coupled plasma atomic emission spectrometry (ICP-AES) and mass spectrometry (ICP-MS). The interference from the glass fiber filters and the variations of the detection limits of the major elements with total suspended particulates (TSP) were thoroughly examined. According to these results, the concentrations of most of the elements in suspended particulates can be determined via the proposed method. On the other hand, determining Na is relatively difficult owing to significant interference from glass fiber filters of the beta-gauge monitoring system. In addition, real samples collected from an air-pollution episode in Taiwan during the spring of 1995 were analyzed, in which the proposed method’s effectiveness was confirmed. The analysis results also indicate that “Kosa” aerosols, having originated in China, might have been the culprit of that episode.

Journal

  • Analytical Sciences

    Analytical Sciences 14 (4), 763-768, 1998

    The Japan Society for Analytical Chemistry

Citations (4)*help

See more

References(30)*help

See more

Report a problem

Back to top