Lead(II)-Selective Electrode Based on Phenyl Disulfide.

DOI PDF Web Site Web Site Web Site ほか1件をすべて表示 一部だけ表示 被引用文献1件 参考文献38件

書誌事項

タイトル別名
  • Lead 2 Selective Electrode Based on Phenyl Disulfide

この論文をさがす

説明

A PVC membrane electrode for lead ions based on phenyl disulfide as the membrane carrier was developed. The electrode exhibits a good Nernstian slope of 29.3 ± 0.7 mV/decade and a linear range of 2.0 × 10-6 - 1.0 × 10-2 M for Pb(NO3)2. The limit of detection is 1.2 × 10-6 M. It has a response time of 45 s and can be used for at least fifty days without any divergence in potential. The proposed membrane sensor revealed high selectivity for Pb2+ over a wide variety of other metal ions and could be used in the pH range of 3.5 - 6.3. The electrode was used as an indicator electrode in potentiometric titration of lead ions.

収録刊行物

  • Analytical Sciences

    Analytical Sciences 18 (9), 987-991, 2002

    社団法人 日本分析化学会

被引用文献 (1)*注記

もっと見る

参考文献 (38)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ