Calibration Method for the Gas-Chromatographic Retention Time of Polychlorinated Biphenyl Congeners.
-
- KATO Hiroshi
- System Development Division, JEOL Ltd.
-
- UEDA Yoshihisa
- Analytical Instrument Division, JEOL Ltd.
-
- NAKATA Munetaka
- Graduate School of Bio-Applications and Systems Engineering, Tokyo University of Agriculture and Technology
この論文をさがす
抄録
A calibration method for the gas-chromatographic (GC) retention time of polychlorinated biphenyl (PCB) congeners is presented. This method calculates retention time of all PCB congeners in the GC system of an analytical laboratory from one reported in the literature with a commercial PCB mixture and some pure congener standards. Two corrections were implemented to compensate for any retention differences arising from the GC condition changes, such as the column dimensions, temperature programs and flow rates. One is an instrument function correction that compensates for common congener behavior. The other is a selectivity correction that compensates for individual congener behavior. For a selectivity correction, new descriptors of a quantitative structure-retention relationship (QSRR) were developed. The method was applied to several stationary phases to evaluate its ability and precision. The standard errors of the calculated retention times were in most cases within a chromatographic peak width.
収録刊行物
-
- Analytical Sciences
-
Analytical Sciences 16 (7), 693-699, 2000
社団法人 日本分析化学会
- Tweet
キーワード
詳細情報 詳細情報について
-
- CRID
- 1390282679233392128
-
- NII論文ID
- 130004440287
- 10005719971
-
- NII書誌ID
- AA10500785
-
- COI
- 1:CAS:528:DC%2BD3cXltlylu74%3D
-
- ISSN
- 13482246
- 09106340
-
- NDL書誌ID
- 5472829
-
- 本文言語コード
- en
-
- データソース種別
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可