Comparison of Enzymatic Recycling Electrodes for Measuring Aminophenol: Development of a Highly Sensitive Natriuretic Peptide Assay System

  • MIE Yasuhiro
    Hokkaido Center, National Institute of Advanced Industrial Science and Technology (AIST)
  • KOWATA Keiko
    Hokkaido Center, National Institute of Advanced Industrial Science and Technology (AIST)
  • HIRANO Yu
    Hokkaido Center, National Institute of Advanced Industrial Science and Technology (AIST)
  • NIWA Osamu
    Tsukuba Center, National Institute of Advanced Industrial Science and Technology (AIST)
  • MIZUTANI Fumio
    Hokkaido Center, National Institute of Advanced Industrial Science and Technology (AIST)

この論文をさがす

抄録

Several redox enzymes were examined for enzymatic/electrochemical-recycling systems in order to measure p-aminophenol (PAP) with high sensitivity. Glucose oxidase (GOD) and diaphorase (DI) worked well as catalysts for recycling electrode systems: these enzymes effectively reduced p-iminoquinone (PIQ), the electrochemically-oxidized form of PAP, and caused an enhancement in the electrochemical signals (anodic currents in the voltammogram and amperogram) by ∼ 100 fold. The lower detection limits for PAP were estimated to be 50 nM with the GOD system and 2 nM with the DI system. We combined the enzymatic-recycling electrode using DI with an enzyme immunoassay system to measure atrial natriuretic peptide (ANP), an important marker peptide hormone involved in heart diseases. ANPs from serum samples at ppt-levels were determined appropriately using the present assay system.

収録刊行物

  • Analytical Sciences

    Analytical Sciences 24 (5), 577-582, 2008

    社団法人 日本分析化学会

被引用文献 (1)*注記

もっと見る

参考文献 (60)*注記

もっと見る

キーワード

詳細情報 詳細情報について

問題の指摘