A Novel Membrane Sensor for Histamine H1-Receptor Antagonist "Fexofenadine"

DOI PDF Web Site Web Site Web Site ほか1件をすべて表示 一部だけ表示 参考文献57件
  • ABBAS M. N.
    Microanal. Lab., Appl. Org. Chem. Dept., National Research Centre
  • ABDEL FATTAH A. A.
    Chem. Dept. Fac. Sci., Benha Branch, Zagazig University
  • ZAHRAN E.
    Microanal. Lab., Appl. Org. Chem. Dept., National Research Centre

書誌事項

タイトル別名
  • Novel Membrane Sensor for Histamine H1 Receptor Antagonist Fexofenadine

この論文をさがす

抄録

The construction and general performance of thirteen new polymeric membrane sensors for the determination of fexofenadine hydrochloride based on its ion exchange with reineckate, tetraphenylborate and tetraiodomercurate have been studied. The effects of membrane composition, type of plasticizer, pH value of sample solution and concentration of the analyte in the sensor internal solution have been thoroughly investigated. The novel sensor based on reineckate exchanger shows a stable, potentiometric response for fexofenadine in the concentration range of 1 × 10-2 - 2.5 × 10-6 M at 25°C that is independent of pH in the range of 2.0 - 4.5. The sensor possesses a Nernstian cationic slope of 62.3 ± 0.7 mV/concentration decade and a lower detection limit of 1.3 × 10-6 M with a fast response time of 20 - 40 s. Selectivity coefficients for a number of interfering ions and excipients relative to fexofenadine were investigated. There is negligible interference from almost all studied cations, anions, and pharmaceutical excipients, however, citrizine that has a structure homologous to that of fexofenadine was found to interfere. The determination of fexofenadine in aqueous solution shows an average recovery of 99.83% with a mean relative standard deviation (RSD) of 0.5%. Direct potentiometric determination of fexofenadine in tablets gave results that compare favorably with those obtained by standard spectrophotometric methods. Potentiometric titration of fexofenadine with phosphomolybdic acid as a titrant has been monitored with the proposed sensor as an end point indicator electrode.

収録刊行物

  • Analytical Sciences

    Analytical Sciences 20 (8), 1137-1142, 2004

    社団法人 日本分析化学会

参考文献 (57)*注記

もっと見る

キーワード

詳細情報 詳細情報について

問題の指摘

ページトップへ