Determination of Trace Impurities in High-Purity Cadmium by High-Resolution Inductively Coupled Plasma Mass Spectrometry

  • XIE Hualin
    Department of Chemistry, Hunan Institute of Technology
  • NIE Xidu
    Department of Chemistry, Hunan Institute of Technology

Search this article

Abstract

Trace impurities in high-purity cadmium were determined by high-resolution inductively coupled plasma mass spectrometry (HR-ICP-MS). To overcome some potentially problematic spectral interference, measurements were acquired in both medium and high-resolution modes. The matrix effects due to the presence of excess HCl and cadmium were evaluated. The optimum conditions for the determination were tested in this experiment and discussed. The detection limits ranged from 0.01 to 0.27 µg g-1, depending on the elements. The results for the determination of 22 trace elements in high-purity cadmium are presented.

Journal

  • Analytical Sciences

    Analytical Sciences 22 (10), 1371-1374, 2006

    The Japan Society for Analytical Chemistry

References(28)*help

See more

Report a problem

Back to top