Coulometric Titration of Urea with Electrogenerated Hypobromite
-
- KATO Jun
- Faculty of Engineering, Tokyo University of Science Nuclear Cycle Backend Directorate, Japan Atomic Energy Agency
-
- KOSEKI Takuma
- Faculty of Engineering, Tokyo University of Science
-
- AOKI Yukie
- Faculty of Engineering, Tokyo University of Science
-
- YAMADA Ayako
- Faculty of Engineering, Tokyo University of Science
-
- TANAKA Tatsuhiko
- Faculty of Engineering, Tokyo University of Science
この論文をさがす
抄録
A definitive method is described for the indirect assay of several tens of milligrams of urea by coulometric titration. Urea was decomposed in concentrated sulfuric acid using a Kjeldahl flask. Subsequently, the formed ammonium ion was titrated with electrogenerated hypobromite ion in a sodium bromide-sodium tetraborate medium of pH 8.6, with amperometric end-point detection. Parameters affecting the pretreatment procedure were evaluated. The optimized conditions included the heating of 2 g of urea at around 300°C for 2 h with 10 cm3 of sulfuric acid. Under the proposed conditions, the assay value with expanded uncertainty (k = 2), 99.870 ± 0.026%, agreed well with the certified value of NIST SRM 912a urea, 99.9 ± 0.1%.
収録刊行物
-
- Analytical Sciences
-
Analytical Sciences 29 (7), 753-755, 2013
社団法人 日本分析化学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390282679235842304
-
- NII論文ID
- 10031185239
-
- NII書誌ID
- AA10500785
-
- COI
- 1:STN:280:DC%2BC3sjpvVarsQ%3D%3D
-
- ISSN
- 13482246
- 09106340
-
- NDL書誌ID
- 024671870
-
- PubMed
- 23842420
-
- 本文言語コード
- en
-
- データソース種別
-
- JaLC
- NDL
- Crossref
- PubMed
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可