Exploiting Zone Trapping to Avoid Liberation of Air Bubbles in Flow-based Analytical Procedures Requiring Heating

DOI PDF Web Site Web Site Web Site ほか1件をすべて表示 一部だけ表示 参考文献15件

この論文をさがす

抄録

In flow-based analytical procedures requiring heating, liberation of air bubbles is avoided by trapping a sample selected portion into a heated hermetic environment. The flow-through cuvette is maintained into a temperature-controlled aluminium block, thus acting as the trapping element and allowing real-time monitoring. The feasibility of the innovation was demonstrated in the spectrophotometric catalytic determination of vanadium in mineral waters. Air bubbles were not released even for temperatures as high as 95°C. The proposed system handles about 25 samples per hour, requires only 3 mg p-anisidine per determination and yields precise results (r.s.d. = 2.1%), in agreement with ICP-MS. Detection limit was evaluated (3.3 σ criterion) as 0.1 μg L−1 V.

収録刊行物

  • Analytical Sciences

    Analytical Sciences 30 (8), 833-838, 2014

    社団法人 日本分析化学会

参考文献 (15)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘