Signal Enhancement by Crossing the Sample Flow at a Small Angle against the Laser Beam in Multiphoton Ionization Mass Spectrometry

DOI PDF Web Site Web Site Web Site ほか1件をすべて表示 一部だけ表示 被引用文献5件 参考文献21件
  • MATSUI Taiki
    Department of Applied Chemistry, Graduate School of Engineering, Kyushu University
  • IMASAKA Totaro
    Department of Applied Chemistry, Graduate School of Engineering, Kyushu University Division of Optoelectronics and Photonics, Center for Future Chemistry, Kyushu University

この論文をさがす

抄録

To enhance sensitivity in multiphoton ionization/time-of-flight mass spectrometry, the sample, in the form of an effusive molecular flow, was introduced at a small angle relative to the laser beam. This approach enlarged the volume of interaction between the sample and the laser beam, thus increasing the number of ions formed by multiphoton ionization. The sensitivity obtained using this technique was compared with that obtained using a previous method based on the off-axis sample introduction technique with detection limits for organic hydrocarbons at the subfemtogram level. The detection limit was examined for dioxins and an improvement of 2.5 ± 0.4 fold was found. The technique developed herein is applicable for practical ultratrace analysis of persistent organic pollutants.

収録刊行物

  • Analytical Sciences

    Analytical Sciences 30 (4), 445-449, 2014