Temperature-responsive Solid-phase Extraction Column for Biological Sample Pretreatment

DOI PDF Web Site Web Site Web Site View 1 Remaining Hide 7 Citations 17 References

Search this article

Abstract

We have developed a novel solid-phase extraction (SPE) system utilizing a temperature-responsive polymer hydrogel-modified stationary phase. Aminopropyl silica beads (average diameter, 40 – 64 μm) were coated with poly(N-isopropylacrylamide) (PNIPAAm)-based thermo-responsive hydrogels. Butyl methacrylate (BMA) and N,N-dimethylaminopropyl acrylamide (DMAPAAm) were used as the hydrophobic and cationic monomers, respectively, and copolymerized with NIPAAm. To evaluate the use of this SPE cartridge for the analysis of drugs and proteins in biological fluids, we studied the separation of phenytoin and theophylline from human serum albumin (HSA) as a model system. The retention of the analytes in an exclusively aqueous eluent could be modulated by changing the temperature and salt content. These results indicated that this temperature-responsive SPE system can be applied to the pretreatment of biological samples for the measurement of serum drug levels.

Journal

  • Analytical Sciences

    Analytical Sciences 31 (9), 881-886, 2015

    The Japan Society for Analytical Chemistry

Citations (7)*help

See more

References(17)*help

See more

Related Projects

See more

Report a problem

Back to top