A Fast Way to Make a Monolithic Column for a High Pressure Electroosmotic Pump

この論文をさがす

抄録

A simple way was proposed to make a monolithic column for a high pressure electroosmotic pump (EOP). It is in-situ synthesized inside the silica capillary from potassium silicate solution and no frit is required. Compared with common approaches to make columns for EOP, the present method is robust and fast (<4 h). For pure water, a stand-alone EOP operated at 15 kV applied voltage is capable of generating a flow rate of 3.1 μL/min and a maximum static pressure of ∼5.4 MPa.

収録刊行物

  • Analytical Sciences

    Analytical Sciences 26 (8), 921-923, 2010

    社団法人 日本分析化学会

参考文献 (18)*注記