フィルムバッジと他の個人線量計による被ばく線量測定値の比較
書誌事項
- タイトル別名
-
- Comparison of Measured Doses by Film Badge and Other Personnel Dosimeters
この論文をさがす
説明
In order to study a discrepancy in external dose measured by various types of personnel dosimeters, especially between a film badge and other dosimeters, the energy and directional dependence of 6 kinds of dosimeters was examined both in free air and on the phantom. On the basis of the experimental results, it was found that the backscattered radiation from the body has stronger influences on the dose by a film badge than that by other types of dosimeters and that the discrepancy is mainly caused by the scattered radiation. Therefore, it is suggested that a film badge should be calibrated on the phantom for diminishing the discrepancy.
収録刊行物
-
- 保健物理
-
保健物理 12 (2), 85-93, 1977
一般社団法人日本保健物理学会