Electrolyte counterflow in isotachophoresis
-
- AKIYAMA Junichi
- Shimadzu Seisakusho Co. Ltd.
-
- MIZUNO Tosie
- Shimadzu Seisakusho Co. Ltd.
Bibliographic Information
- Other Title
-
- 細管式等連電気泳動法におけるカウンター・フローの効果
- サイカンシキ トウソク デンキ エイドウホウ ニ オケル カウンター フロー
- 細管式等速電気泳動法におけるカウンター・フローの効果
Search this article
Description
A syring-type counterflow device for isotachophoresis has been developed to separate ions that differ only little in mobility. The device was installed in a Shimadzu Isotachophoretic Analyzer IP-1, in order to study the effects of the counterflow on the separation efficiency. The migration current was set to 100 μA; the capillary tube was 50 cm long; the sample was a mixture of aqueous solutions of 0.01 M sodium oxalate, 0.01M sodium formate and 0.01M sodium citrate. The rate of counterflow was varied in 6 steps of 6, 12, 25, 50, 100 and 200 μl/h, by changing the plunger speed. Thus, separation efficiency has been studied as a function of sample volume and counterflow rate. A mixed zone of formate and citrate was observed for a sample volume above 8 μl when counterflow was not applied. This mixed zone was eliminated when counterflow was applied at 200μl/h for 15 min. The effective length of the capillary tube can be increased by 40% by applying a counterflow.
Journal
-
- BUNSEKI KAGAKU
-
BUNSEKI KAGAKU 24 (11), 728-729, 1975
The Japan Society for Analytical Chemistry
- Tweet
Details 詳細情報について
-
- CRID
- 1390282679331753088
-
- NII Article ID
- 110002907492
-
- NII Book ID
- AN00222633
-
- NDL BIB ID
- 1629240
-
- ISSN
- 05251931
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- NDL
- Crossref
- NDL-Digital
- CiNii Articles
-
- Abstract License Flag
- Disallowed