書誌事項
- タイトル別名
-
- Two-dimensional Finite Element Stress Analysis on the Effect of Bending Position of Blade Type Dental Implant.
この論文をさがす
抄録
This study was to investigate the stress distribution in the peripheral bone tissue of the post of the blade type implant under changing various bending positions by using two-dimensional finite element method analysis and the following conclusions were obtained: 1) As the bending position of the post of the implant being away from the cortical bone surface, maximum principal compressive strength value and its range were decreased. 2) In the relation of the implant body and its peripheral supporting tissue, stresses concentrated at the cortical bone in the bone contact condition. 3) In the relation of the implant body and its peripheral supporting tissue, stresses concentrated at the fibrosis tissue intermediating condition. Moreover these stresses spreaded from the cortical bone to the cancellous bone area. Therefore the design which must be bent the post of the implant, the bending portion should be away from the cortical bone surface.
収録刊行物
-
- 日本補綴歯科学会雑誌
-
日本補綴歯科学会雑誌 37 (5), 951-960, 1993
社団法人 日本補綴歯科学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390282679385707008
-
- NII論文ID
- 110003856456
-
- NII書誌ID
- AN00197911
-
- ISSN
- 1883177X
- 03895386
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可