Successful Repair Using VALIANT Captivia for Type 1 Endoleak through a Fenestration on the Najuta Endograft in the Treatment of Distal Arch Aneurysm

Bibliographic Information

Other Title
  • Najutaプロトタイプモデルによる遠位弓部大動脈瘤治療後の開窓部エンドリークに対しVALIANT Captiviaの内挿が有効であった一例

Search this article

Description

<p>Endograft failure, especially type 1 endoleak, has remained a significant clinical concern and its treatment is often challenging. We report a case of type 1 endoleak after TEVAR using the Najuta prototype, which is a semi-custom-made pre-curved fenestrated device. The endoleak was successfully repaired by additional TEVAR using VALIANT Captivia. An 85-year old man with 70% of Japan Score (estimated early mortality rate) was admitted to our hospital owing to type 1 endoleak and enlargement of a distal arch aneurysm. He underwent zone 0 TEVAR using the Najuta prototype for the treatment of distal arch aneurysm at age 80. Computed tomography showed type 1 endoleak through a fenestration on the stent graft (SG) at age 83. So we performed re-intervention using VALIANT Captivia. This SG has proximal bare stent and strong radial force that may be effective for adequate seal of a space around a fenestration of the Najuta endograft. After this procedure, the endoleak disappeared.</p>

Journal

Details 詳細情報について

Report a problem

Back to top