Properties of All-solid-state pH Sensor Using LaF3 Single Crystal
-
- KUWATA Shigeki
- Department of Applied Chemistry and Biotechnology, Niihama National College of Technology
-
- NAKAYAMA Susumu
- Department of Applied Chemistry and Biotechnology, Niihama National College of Technology
Bibliographic Information
- Other Title
-
- LaF3単結晶を用いた全固体型pHセンサの特性
Search this article
Description
An all-solid-state type pH sensor was investigated for the monitoring of pH in the aqueous solution in the tempera-ture range from 25°C to 80°C. The all-solid-state pH sensor consists of two half-cells: an indicator electrode using the LaF3 single crystal and an Ag/AgCl reference electrode coated with Nafion film. The electromotive force of the all-solid-state pH sensor decreased linearly with pH increase in the solution in accordance with the Nernst's equation. The electromotive force was also scarcely influenced in the presence of inorganic ions such as Na+ and Cl-. It was practi-cally confirmed by the pH titration test that the all-solid-state pH sensor behaved similarly to the commercial pH meter with the conventional glass electrode. In addition, the all-solid-state pH sensor operated stably up to 80°C.
Journal
-
- Journal of Ion Exchange
-
Journal of Ion Exchange 19 (3), 174-178, 2008
Japan Society of Ion Exchange
- Tweet
Keywords
Details 詳細情報について
-
- CRID
- 1390282679411162240
-
- NII Article ID
- 130004162506
-
- ISSN
- 18843360
- 0915860X
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- Crossref
- CiNii Articles
- OpenAIRE
-
- Abstract License Flag
- Disallowed