人工股関節ステム形状の最適設計

書誌事項

タイトル別名
  • Shape Optimization of an Artificial Hip Prosthesis.
  • ジンコウ コ カンセツ ステム ケイジョウ ノ サイテキ セッケイ

この論文をさがす

抄録

When cemented hip replacement implants are developed using a finite element method (FEM), the validation of the inplants comparing with experimental tests and clinical survival data should be required. As the first step, the optimum shape of the femoral component of total hip arthroplasty (THA) was designed to minimize the stress peak in the cement mantle that is the primal cause for fixation failure. In the next step, the strain in the cement mantle surrounding the cemented femoral components of THA was measured using strain gauges embedded within the cement mantle adjacent to the developed femoral stem to validate the optimization results of FEM. As a result of this study, optimum shape of femoral component was developed and the FEM results were validated by strain gauge experiments.