A high sensitivity CMOS bio photosensor applicable to a bioluminescent assay was developed with a 0.18µm CMOS image sensor (CIS) process. The bio photosensor consisting of a photosensor and a PWM 20bit A/D converter achieved high sensitivity for detecting a extremely low bioluminescence due to a large photodiode area, a long exposure time and the other noise reduction techniques. The bio photosensor chip has a 2×4 sensor array on a 2.45×2.45mm2 die. Experimental results with the bioluminescence showed the chip can detect below 10-5lux luminescence at room temperature and the power consumption is 32µW.