Measurement and Stability of Range Ratios of Standard Radiation Thermometers
-
- SAKUMA Fumihiro
- National Institute of Advanced Industrial Science and Technology
-
- MA Laina
- National Institute of Advanced Industrial Science and Technology
Bibliographic Information
- Other Title
-
- 標準放射温度計のレンジ比の測定とその安定性
- ヒョウジュン ホウシャ オンドケイ ノ レンジヒ ノ ソクテイ ト ソノ アンテイセイ
Search this article
Abstract
0.65μm and 0.9μm radiation thermometers are used as the standard thermometers at high temperatures. These radiation thermometers employ three to five ranges to cover their whole temperature ranges. The range ratios of these thermometers influence the calibration uncertainty. The measurement method, measured results and the long term stability of the range ratios are shown.
Journal
-
- Transactions of the Society of Instrument and Control Engineers
-
Transactions of the Society of Instrument and Control Engineers 44 (3), 209-218, 2008
The Society of Instrument and Control Engineers
- Tweet
Details 詳細情報について
-
- CRID
- 1390282679480493696
-
- NII Article ID
- 10021990059
-
- NII Book ID
- AN00072392
-
- ISSN
- 18838189
- 04534654
-
- NDL BIB ID
- 9437075
-
- Data Source
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- Abstract License Flag
- Disallowed