New supplemental activator for lead isotope analysis using TIMS
-
- NOHDA SUSUMU
- Earth Dynamic System Research Center, National Cheng Kung University
-
- WANG BO-SHIAN
- Earth Dynamic System Research Center, National Cheng Kung University
-
- YOU CHEN-FENG
- Earth Dynamic System Research Center, National Cheng Kung University
-
- SHINJO RYUICHI
- Department of Physics and Earth Sciences, University of the Ryukyus
この論文をさがす
抄録
We have devised a new supplementary activator for lead isotope measurements performed using thermal ionization mass spectrometry (TIMS). The activator consisted of a mixture of 250 ppm Ge and 500 ppm Re and was used with normal colloidal silica. This new substance shows excellent performance in terms of enhancing the ion beam intensity. The supplementary activator yielded 3.5–4.5 × 10–11 A 208Pb for 10 ng Pb, 1.5–2.5 × 10–11 A for 5 ng Pb, and ~1.5 × 10–11 A for 2.5 ng Pb samples, indicating that the new activator was superior to those currently in use. This improvement should be applicable to samples with low Pb content, for example, in environmental tracer studies and Pb–Pb chronology studies.
収録刊行物
-
- GEOCHEMICAL JOURNAL
-
GEOCHEMICAL JOURNAL 45 (2), 169-174, 2011
一般社団法人日本地球化学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390282679529732608
-
- NII論文ID
- 10028198583
-
- NII書誌ID
- AA00654975
-
- COI
- 1:CAS:528:DC%2BC3MXnsFyksr8%3D
-
- ISSN
- 18805973
- 00167002
-
- NDL書誌ID
- 11047244
-
- 本文言語コード
- en
-
- データソース種別
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-