ディジタルフィルタを用いた光電脈波法による心拍数・呼吸数同時モニタ
書誌事項
- タイトル別名
-
- Photoplethysmography for Simultaneous Monitoring of Heart and Respiratory Rates Using Digital Filters.
この論文をさがす
説明
A monitor for simultaneous recording of heart and respiratory rates has been developed using digital filters and photoplethysmography (PPG) with plastic optical fiber. PPG signal includes both heart and respiratory synchronous components. Digital filtering technique was employed to differentiate between those. With increasing heart and respiratory rates, the cut-off frequency of filter was changed empirically. Finite impulse response digital filters were designed using the Remez exchange algorithm. One filter for the heart-related signal was 40th order of Chebyshev type and the other filters for the respiration-related signal were 60th order of Chebyshev type. A suitable cut-off frequency of filter was selected empirically. The heart-related signal was detected by using a impulse response, while the respiration-related signal was detected by three impulse responses. One out of three impulse response was selected for the respiration-related signal. Heart and respiratory rate were increased with whole body heart stress i. e. bathing. The heart rates from 65 to 87 beats/min and the respiratory rates from 13 to 27 breath/min agreed with those obtained with ECG and thermistor. The digital filtering technique may provide on-line monitor for heart and respiratory rates.
収録刊行物
-
- 医用電子と生体工学
-
医用電子と生体工学 31 (4), 360-366, 1993
一般社団法人 日本生体医工学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390282679531793408
-
- NII論文ID
- 130004326879
-
- ISSN
- 21855498
- 00213292
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可