STUDY ON COD-ESTIMATION IN RIVER AND WASTE WATER BY PORTABLE ULTRAVIOLET ABSORBANCE PHOTOMETR

Bibliographic Information

Other Title
  • 発展途上国を対象とした簡易UV計による有機物汚濁測定法

Search this article

Description

A simple, inexpensive and portable ultraviolet absorbance photometer was developed for COD estimation not only for river water but also for waste water. The performance of this instrument such as Zero-drift, 100%-drift were very good. Moreover good liner response with slope 1 was obtained between standard solution (potassium hydrogen phthalete) and absorbance at ultraviolet 254 nm. Using filtered various testing solutions with 0.45μm pore size (river water, leachate, food industry, animal farm and septic tank), we compared the absorbance from this instrument with conventional spectrophotometer. Finally, good linear relationship: UV254 (this instrument) =0.98×UV254 (spectrophotometer) -0.004 (R=0.99) was obtained. For the samples collected in the Japan, COD-Mn was estimated by following equation: COD-Mn (mg/l) =60×UV254+1.3, and for samples collected in Japan, the Philippines and Thailand, <BR>COD-Cr (mg/l) =169×UV254. However, poor results were obtained from testing solutions which contained high concentration of SS (>30mg/l ), colored or fluorescent brightening agent. Although, the weak points for these samples were overcome by a simple operation, such as dilution.

Journal

Citations (1)*help

See more

Details 詳細情報について

  • CRID
    1390282679567230848
  • NII Article ID
    130004173810
  • DOI
    10.5956/jriet.32.575
  • ISSN
    18828590
    03889459
  • Text Lang
    ja
  • Data Source
    • JaLC
    • Crossref
    • CiNii Articles
  • Abstract License Flag
    Disallowed

Report a problem